Operation Liquiline System CA80SI
86 Endress+Hauser
Function = Diagnostics
Function Options Info
Operating mode Selection
• as assigned
• Namur M
• Namur S
• Namur C
• Namur F
Factory setting
as assigned
as assigned
If this option is selected, the diagnostic messages
which you have individually assigned to the relay
are output via the relay.
Namur M ... Namur F
If you decided to use one of the Namur classes,
all the messages that are assigned to the
individual class are output via the relay. You can
also change the Namur class assignment for
every diagnostic message.
(Menu/Setup/General settings/Extended
setup/Diagnostics settings/Diag. behavior or
Menu/Setup/Inputs/<Sensor>/Extended
setup/Diagnostics settings/Diag. behavior)
Attributed diagnostic messages
Operating mode = as
assigned
Read only All the messages assigned to the relay output
appear on the display. You do not have the
option of editing the information here.
Outputting the status of a cleaning function
Function = Cleaning (for sensors)
Function Options Info
Assignments Options
• None
• Depends on the type of
cleaning
Factory setting
None
Here you can specify how a cleaning function
should be displayed for the relay.
Depending on the cleaning program selected
(Menu/Setup/Additional functions/Cleaning)
you can choose from the following:
• Cleaning type = Standard clean
Cleaning 1 - Water, Cleaning 2 - Water,
Cleaning 3 - Water, Cleaning 4 - Water
• Cleaning type = Chemoclean
Cleaning 1 - Water, Cleaning 1 - Cleaner,
Cleaning 2 - Water, Cleaning 2 - Cleaner,
Cleaning 3 - Water, Cleaning 3 - Cleaner,
Cleaning 4 - Water, Cleaning 4 - Cleaner
• Cleaning type = Chemoclean Plus
4x Cleaning 1 - %0V, 4x Cleaning 2 - %0V
1)
1) %0V is variable text which you can assign in Menu/Setup/Additional functions/Cleaning/Chemoclean
Plus/Output label 1 ... 4 .
Outputting the "Measurement active" system status and the information "Sample
required"
Function = Analyzer
Function Options Info
Signal type = Static signal
Assignments
Function = Analyzer
Selection
• None
• Sample required SP1
• Measurement active SP1
Factory setting
None
If this option is selected, information is output as
to whether a measurement is active on the
selected measuring channel or an action is
started that requires sample (measurement,
calibration or cleaning).