Operation Liquiline System CA80SI
90 Endress+Hauser
Source of data Measured value
Controller 1
Current input 1 ... 3
Selection
Bipolar
(only for current outputs)
Unipolar+
Unipolar-
Controller 2
Temperature 1 ... 3
Mathematical functions All the mathematical functions can also be used as a data source and the calculated
value can be used as the measured value.
You can monitor the manipulated variable by assigning the controller manipulated
variable to a limit switch (e. g. configure a dosing time alarm).
Menu/Setup/Additional functions/Limit switches/Limit switch 1 ... 8
Function Options Info
Cleaning program Selection
None
Cleaning 1 ... 4
Factory setting
None
Use this function to choose which cleaning
instance should start when the limit switch is
active.
Function Selection
Off
On
Factory setting
Off
Activating/deactivating the limit switch
Operating mode Selection
Above limit check
Below limit check
In range check
Out of range check
Change rate
Factory setting
Above limit check
Type of limit value monitoring:
Limit value overshoot or undershoot→  67
Measured value within or outside a range
→  68
Rate of change →  70
Limit value Settings depend on the
measured value
Operating mode = Above limit check or Below
limit check
1
2
t
MV
t
1
t
2
t
3
t
4
t
5
1
2
t
MV
t
1
t
2
t
3
t
4
t
5
A B
A0028523
 67 Exceeding (A) and undershooting (B) a limit value (without hysteresis and switch-on delay)
1 Limit value
2 Alarm range
t
1,3,5
No action
t
2,4
An event is generated
If the measured values (MV) are increasing, the relay contact is closed when the on-value is exceeded (Limit
value + Hysteresis) and the start delay (Start delay) has elapsed.
If the measured values are decreasing, the relay contact is reset when the off-value is undershot (Limit value
- Hysteresis) and following the drop-out delay (Switch off delay).