FaroArm
®
Manual
January 2009
80
Chapter 2: CAM2® Measure Devices Menu
Single Hole Method - Guidance
Perform the Single Hole calibration using the FARO probe calibration
cone. If the G
UIDANCE check box is selected for the current probe, use
the following steps. See “Edit Probe” on page 78.
Collect all of the points in this method by holding down the FRONT
button. The FaroArm will collect points as fast as possible (“scanning”)
until you release the FRONT button. Digitize at least 200 points in each
of the three different positions.
Figure 2-5 Single Hole Method
Position #1
Position #2
Position #3
08M46E00_FaroArmUSB.book Page 80 Tuesday, February 17, 2009 1:42 PM