FLUXUS F808, F809 6 Installation of FLUXUS F808
50 UMFLUXUS_F808_8091V1-2-1EN, 2019-08-23
active current
loop/HART
F808**-A1
F808**-F1
current loop R
ext
< 500 Ω
U
int
= 24 V
I1: 2 (+)
KL3
I1: 1 (-)
HART mode
I1: 2 (+)
KL3
I1: 1 (-)
passive current
loop/HART
F808**-A1
F808**-F1
current loop U
ext
=
7...30 V
U
ext
> 0.022 A
.
R
ext
[Ω] + 7 V
example:
U
ext
= 12 V
R
ext
≤ 227 Ω
current during transmitter error:
I
fault
= 3.2...3.5 mA
I1: 1 (-)
KL3
I1: 2 (+)
HART mode
I1: 1 (-)
KL3
I1: 2 (+)
binary output
(open collector) B1: 6 (+)
KL1
B1: 5 (-)
U
ext
= 5...24 V
R
c
[kΩ] = U
ext
/I
c
[mA]
I
c
= 1...4 mA
binary output
(open collector) B1: 6 (+)
KL1
B1: 5 (-)
U
ext
= 5...30 V
R
c
[kΩ] = U
ext
/I
c
[mA]
I
c
= 2...100 mA
I
off
≤ 0.4 mA
Tab. 6.20: Circuits of the outputs
output transmitter
external circuit
remark
internal circuit connection
The number, type and connections of the outputs are customized.
R
ext
is the sum of all ohmic resistances in the circuit (e.g. resistance of the conductors, resistance of the amperemeter/voltmeter).
+
-
mA
+
-
ƒ
G
U
int
+
-
mA
mA
U
ext
-
+
ƒ
G
ƒ
G
mA
U
ext
-
+
R
c
U
ext
V
+
-
R
c
U
ext
V
+
-

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F808 A1 Series and is the answer not in the manual?

Flexim FLUXUS F808 A1 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F808 A1 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals