FLUXUS F808, F809 19 Maintenance
162 UMFLUXUS_F808_8091V1-2-1EN, 2019-08-23
19 Maintenance
The transmitter and the transducers are practically maintenance-free.
19.1 Cleaning
Clean the transmitter with a soft cloth. Do not use detergents.
Remove traces of the coupling compound from the transducers with a soft paper towel.
19.2 Examination of the O-ring
Check the O-ring when opening the housing (FLUXUS F808, see Fig. 19.1, FLUXUS F809, see Fig. 19.2).
Flameproof enclosure housing
If the O-ring is defective or the housing does not seal anymore, change it (see Tab. 19.1).
Housing with increased safety
If the O-ring is defective or the housing does not seal anymore, it has to be replaced.
Attention! O-rings on increased safety housings may only be replaced by trained FLEXIM personnel.
Note! Store the O-ring replacement in a hermetically sealed and dark place.
Fig. 19.1: FLUXUS F808
Fig. 19.2: FLUXUS F809
flameproof enclosure
housing
housing
with increased safety
O-ring
O-ring
flameproof enclosure
housing
housing
with increased safety
housing with increased
safety
O-ring
O-ring
O-ring

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F808 A1 Series and is the answer not in the manual?

Flexim FLUXUS F808 A1 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F808 A1 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals