FLUXUS F808, F809 15 Settings
138 UMFLUXUS_F808_8091V1-2-1EN, 2019-08-23
15 Settings
15.1 Time and date
The transmitter has a battery-powered clock. Measured values are automatically stored with date and time.
15.1.1 Time
15.1.2 Date
After the time has been set, DATE is displayed.
15.2 Dialogs and menus
15.2.1 Pipe circumference
Select
Special Funct.\SYSTEM settings
\
Set Clock. Press ENTER.
The current time is displayed. Select ok to confirm the time or new to set the time.
Press ENTER.
Press key to select the digit to be edited.
Press key to edit the selected digit. Press ENTER.
The new time is displayed. Select ok to confirm the time or new to set the time again.
Press ENTER.
Select ok to confirm the date or new to set the date. Press ENTER.
Press key to select the digit to be edited.
Press key to edit the selected digit. Press ENTER.
The new date is displayed. Select ok to confirm the date or new to set the date again.
Press ENTER.
Select Special Funct.\SYSTEM settings\Dialogs/Menus. Press ENTER.
Note! The settings of the menu item Dialogs/Menus will be stored at the end of the dialog. If the menu
item is quit before the end of the dialog, the settings will not be stored.
Select on if the pipe circumference is to be entered instead of the pipe diameter in the pro-
gram branch Parameter. Press ENTER.
If on has been selected for Pipe Circumfer., the outer pipe diameter will nevertheless
be requested in the program branch Parameter.
To select the menu item Pipe Circumfer., enter 0 (zero). Press ENTER.
The value displayed in Pipe Circumfer. is calculated on the basis of the last displayed
value of the outer pipe diameter.
example: 100 mm
.
π = 314.2 mm
Enter the pipe circumference. The limits for the pipe circumference are calculated on the
basis of the limits for the outer pipe diameter.
SYSTEM settings↕
Set Clock
TIME 11:00
ok >NEW<
TIME 11:00
Set Time !
TIME 11:11
>OK< new
DATE 2011-01-25
ok >NEW<
DATE 2011-01-25
Set Date !
DATE 2011-01-26
>OK< new
SYSTEM settings↕
Dialogs/Menus
Pipe Circumfer.
off >ON<
Outer Diameter
100.0 mm
Pipe Circumfer.
314.2 mm
Pipe Circumfer.
180 mm

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F808 A1 Series and is the answer not in the manual?

Flexim FLUXUS F808 A1 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F808 A1 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals