13 User modes
13.1 StandardUser mode FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
102
User mode selection
• Select the menu item User mode in the program branch Special functions\System settings.
• Press ENTER.
•Select StandardUser.
• Press ENTER.
Configuration of the settings
• Select the measuring channel in the program branch Options.
• Press ENTER.
• Select the menu item Special settings.
• Press ENTER.
The menu items of the user mode are consecutively displayed.
13.1 StandardUser mode
In the Standarduser mode, all measurements can be carried out for the corresponding application. At the first start-up the
transmitter operates in the StandardUser mode.
Cut-off flow
The cut-off flow is a lower limit for the flow velocity. All measured flow velocities that are below the limit are set to zero.
The cut-off flow can depend on the flow direction.
• Select the menu item Special settings in the program branch Options.
• Press ENTER until the menu item Cut-off flow is displayed.
•Select Off if no value is to be entered for the cut-off flow.
•Select Default if no user-defined inputs are to be made and the default values of ±25 mm/s are to be used.
•Select User-defined to define the cut-off flow value.
• Press ENTER.
All positive values of the flow velocity smaller than this limit are set to zero.
• Enter the cut-off flow.
• Press ENTER.
All negative values of the flow velocity greater than this limit will be set to zero.
• Enter the cut-off flow.
• Press ENTER.
Special functions\System settings\User mode
Options\Special settings
Options\Special settings\Cut-off flow
Options\Special settings\+Cut-off flow
Options\Special settings\-Cut-off flow