15 Data logger
FLUXUS WD 15.1 Configuration of the data logger
117
UMFLUXUS_WDV1-3EN, 2018-07-16
15 Data logger
The transmitter has a data logger which stores the mea-
sured values during the measurement.
The following data is stored within the data logger:
•date
•time
• measuring point number
• pipe parameters
• fluid parameter
• transducer data
• physical quantity
• unit of measurement
• measured value
Measured values transmitted via the output, see section
9.2.5, are also stored in the data logger.
If the pulse values are transmitted via an output, the corre-
sponding flow quantity and the totalizer value are stored in
the data logger. In case of absolute pulse values, the val-
ues of both totalizers are stored.
15.1 Configuration of the data logger
Activation of the channels for storing
• Select the menu item Configuration in the program branch Special functions\Data logger.
• Press ENTER.
• Activate the channels whose measured values have to be stored.
• Press key or to select a channel.
• Press key to activate/deactivate a channel.
• Press ENTER.
Special functions
Data logger
Configuration
Logger channels
Start storing
Storage rate FastF
Ringbuffer
On Off
Storage mode
Average Sample
Store totalizers
Store diagnost. values
Store transd. temp.
Beep on storage
see annex A, p. 154
In order to store measured data, the data logger has
to be configured.
Special functions\Data logger\Configuration\Logger channels