9 Measurement
FLUXUS WD 9.2 Measurement settings
79
UMFLUXUS_WDV1-3EN, 2018-07-16
9.2.5 Configuration of the output
If the transmitter is equipped with outputs, they have to be
configured. In principle, the measured value, the status val-
ue or an event value can be transmitted via the different
outputs.
In the following, the configuration of an analog outputs is
described.
Select the program branch Options.
•Press ENTER.
Selection of the measuring channel
Select the channel (here: Channel A).
•Press ENTER.
This display will not be indicated if the transmitter has only
one measuring channel.
Assignment of an output
•Select Outputs.
Press ENTER.
Select an output to be assigned to the channel (here: Current I1(--)).
Press ENTER.
The scroll list contains all available outputs of the transmitter:
Current Ix (--)
Binary Bx (--)
If the output has already been assigned to a channel, it is displayed as follows: Current I1 (A).
•Select Yes to change the settings for an already assigned output or to assign a new output.
•Select No to cancel the assignment and to return to the previous menu item.
Press ENTER.
Assignment of a source item
One source item has to be assigned to each selected output.
Select a source item whose measured value, status value or event value is to be transmitted via the output, see Tab. 9.1.
Press ENTER.
Options
Channel x
Outputs
Select the output
Enable the output
Source item
see annex A, p. 156
Options\Channel A
Options\Channel A\Outputs\Current I1(--)
Options\Channel A\Outputs\Current I1\I1 Enable
Options\Channel A\Outputs\...\Source item