13 User modes
FLUXUS WD
101
UMFLUXUS_WDV1-3EN, 2018-07-16
13 User modes
The user modes enable an advanced diagnostic of signal
and the measured values as well as the definition of addi-
tional parameters adapted to the application.
The following user modes can be selected:
•StandardUser
•ExpertUser
•SuperUser
•SuperUser ext.
Depending on the selected user mode, different menu items are displayed in the program branch Options\Channel\
Special settings, see Tab. 13.1.
Special functions
System settings
User mode
StandardUser
ExpertUser
SuperUser
SuperUser ext.
see annex A, p. 157
Tab. 13.1: Menu items of the user modes
Options\Special settings StandardUser ExpertUser SuperUser
SuperUser
ext.
default value
Cut-off flow xxxxOn
Flow velocity limit xxx Off
PIG detection xx
Turbulence mode xxxxOff
Max. amplification xxx Off
Pipe signal detection xxx On
LWT pipe wall calibr. xx Off
Linear calibration xx Off
Profile correction xxx Off
Weighting factor xx Off
Multi-point calibration
(if enabled in Special functions\
Measurement\Measurement settings)
xxxx
Start in meas. mode xxxx
Transducer temp. and Transd. temp.
violat. (as source item
Diagnostic values)
xx
Extended diagnost.
(in the program branch Options, calculation
channel)
xxx