9 Measurement
FLUXUS WD 9.1 Parameter input
73
UMFLUXUS_WDV1-3EN, 2018-07-16
9 Measurement
9.1 Parameter input
The pipe and fluid parameters are entered for the selected
measuring point. The parameter ranges are limited by the
technical characteristics of the transducers and the trans-
mitter.
• Select the program branch Parameters.
•Press ENTER.
• Select the channel for which the parameters are to be
entered (here: Channel A).
•Press ENTER.
The display will not be indicated if the transmitter has only
one measuring channel.
9.1.1 Transducer selection
• The transducer (here: CDP2E52) connected to the transmitter is displayed.
• Press ENTER.
This display will only be indicated if transducers and sensor module are connected to the transmitter.
The parameter input can also be carried out without connected transducers or sensor module.
•Select Connected transd.
• Press ENTER.
The display Transducer not found will be indicated if neither transducers nor sensor module are connected.
• Press ENTER.
All settings changed in this program branch will be stored at the beginning of the measurement.
Avoid a simultaneous parameter entry via the keyboard and the USB, Ethernet or process interface. The parameter
records received via these interfaces will overwrite the current transmitter parameterization.
Parameters
Channel x
Transducer
Connected transducer Select transducer
Outer diameter
Pipe material
Pipe wall thickness
Lining
Roughness
Fluid
Fluid temp.
Extension cable
see annex A, p. 152
Parameters\Channel A
The transducer is selected in dependence of the inner
pipe diameter (= outer pipe diameter - 2 × pipe wall
thickness), see Technical specification.
Parameters\Clamp-on transducer CDP2E52
Parameters\Connected transd.