14.2 Activation of a binary output as pulse output
A pulse output is an integrating output which emits a pulse when the volume or the mass of the fluid which has passed the
measuring point, reaches a certain value (pulse value).
The integrated quantity is the selected physical quantity. The integration is restarted as soon as the pulse is emitted. Be-
fore the activation, the binary output has to be configured.
Select the program branch Options.
Press ENTER.
•Select Pulse as source item.
Press ENTER.
Select a list item (here: Pulse +V).
Press ENTER.
Enter the pulse value.
The unit of measurement will be displayed according to the actual physical quantity.
When the counted physical quantity reaches the entered pulse value, a pulse will be transmitted.
Enter the pulse width.
The range of possible pulse widths depends on the specification of the instrument (e.g., counter, PLC) that is to be con-
nected to the output.
Notice!
The activation of a binary output as pulse output is only possible for binary outputs supporting a pulse output.
Options\Channel A\Outputs\Binary B1\Source item\Pulse
Options\Channel A\Outputs\Binary B1\Source item\Pulse\Pulse +V
Options\Channel A\Outputs\Binary B1\Source item\Pulse\Pulse value
Options\Channel A\Outputs\Binary B1\Source item\Pulse\Pulse width