17 Advanced functions
FLUXUS WD 17.1 Totalizers
125
UMFLUXUS_WDV1-3EN, 2018-07-16
17 Advanced functions
17.1 Totalizers
The total volume or total mass of the fluid at the measuring point can be determined.
There are 2 totalizers, one for the positive and the other for the negative flow direction. The unit of measurement used for
totalizing corresponds to the volume or mass unit selected for the physical quantity.
The totalizer values can be displayed in the status line during the measurement, see section 9.7.
17.1.1 Number of decimal places
The values of the totalizers can be displayed with up to 11 places, e.g., 74890046.03. The number of decimal places
(max. 4) can be defined.
• Select the menu item Totalizers in the program branch Special functions.
• Press ENTER.
•Select Automatic if the number of decimal places has to be adjusted automatically.
• Press ENTER.
Low totalizer values will initially be displayed with 3 decimal places. If the values of the totalizers are higher, the number of
decimal places will be reduced.
• Select a value for the number of decimal places.
• Press ENTER.
Tab. 17.1: Key functions
totalizer display press key during the measurement
freezing of the displayed totalizer value press key for at least 2 s during the measurement
display of the flow totalizer for the positive flow direction press key during the measurement
display of the flow totalizer for the negative flow direction press key during the measurement
reset of the totalizers to zero press key 3 times during the measurement
totalizing will be restarted after pressing the key
press key 3 times during the measurement
totalizing will be restarted and displayed immediately
The pressing of a key will only influence the flow totalizers of the measuring channel whose measured values are
currently displayed.
Special functions\Totalizers
max. value display
< 10
6
±0.000 ... ±999999.999
< 10
7
±1000000.00 ... ±9999999.99
< 10
8
±10000000.0 ... ±99999999.9
< 10
10
±1000000000 ... ±9999999999