14 Outputs
FLUXUS WD 14.1 Configuration of a binary output
115
UMFLUXUS_WDV1-3EN, 2018-07-16
If no measurement is carried out, all binary outputs are de-energized, independent of the programmed switching condition.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing key or further information is displayed.
Press ENTER.
If the transmitter possesses a switchable current output, it is displayed whether it is active or passive.
Output function test
The function of the output can now be tested.
Connect a multimeter to the output.
•Select Yes to test the output. Select No to display the next menu item.
Press ENTER.
•Select On to test the energized state of the output.
Press ENTER.
The value has to be low ohmic.
•Select Off to test the de-energized state of the output.
Press ENTER.
The value has to be high ohmic.
•Select Repeat to repeat the test, Finish to display the next menu item.
Press ENTER.
.
•Select Yes to test the status of the output signal. Select No to display the next menu item.
Press ENTER.
Select in the scroll list Passive (idle state) or Active.
Press ENTER.
Depending on the selected switching condition, the value has to be either high or low ohmic.
•Select Repeat to repeat the test, Finish to display the next menu item.
Press ENTER.
Tab. 14.3: Selection of the switching function
property setting description
switching function NO The binary output is energized if the switching condition is met
(event occurs) and de-energized if the switching condition is not
met (idle state).
NC The binary output is de-energized if the switching condition is
met (event occurs) and energized if the switching condition is
not met (idle state).
Options\Channel A\Outputs\...\Output info
Options\Channel A\Outputs\...\Test signal B1
Options\Channel A\Outputs\...\Enter test value B1
Options\Channel A\Outputs\...\Test mea. range B1
Options\Outputs\...\Enter test value B1