14 Outputs
14.1 Configuration of a binary output FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
114
Output function test
The function of the output can now be tested.
• Connect a multimeter to the output.
•Select Yes to test the output. Select No to display the next menu item.
• Press ENTER.
•Select On to test the energized state of the output.
• Press ENTER.
The value has to be low ohmic.
•Select Off to test the de-energized state of the output.
• Press ENTER.
The value has to be high ohmic.
•Select Repeat to repeat the test, Finish to display the next menu item.
• Press ENTER.
.
•Select Yes to test the status of the output signal. Select No to display the next menu item.
• Press ENTER.
• Select in the scroll list Status OK or Status error.
• Press ENTER.
Depending on the selected switching condition, the value has to be either high or low ohmic.
•Select Repeat to repeat the test, Finish to display the next menu item.
• Press ENTER.
14.1.2 Output of the event value
Switching function (idle state)
• Select the switching function, see Tab. 14.3.
• Press ENTER.
Options\Channel A\Outputs\...\Test signal B1
Options\Channel A\Outputs\...\Enter test value B1
Options\Channel A\Outputs\...\Test mea. range B1
Options\Outputs\...\Enter test value B1
Source item
Switching function
(Idle state)
Output info
Test signal
Test mea. range
see annex A, p. 156
Options\Channel A\Outputs\...\Idle state