14 Outputs
FLUXUS WD 14.1 Configuration of a binary output
113
UMFLUXUS_WDV1-3EN, 2018-07-16
• Select the source item to be transmitted to the output by the channel.
14.1.1 Output of a status value
Switching condition (status OK)
• Select the switching function, see Tab. 14.2.
•Press ENTER.
.
If no measurement is carried out, all binary outputs are de-energized, independent of the programmed switching condition.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing key or further information is displayed.
• Press ENTER.
If the transmitter possesses a switchable current output, it is displayed whether it is active or passive.
Tab. 14.1: Output of status values or event values
source item status value event value
physical quantities Flow quantities x
Totalizers x
Fluid properties x
Pulse x
Sound speed x
results Event trigger x
Source item
Switching function
(Status OK)
Output info
Test signal
Test mea. range
see annex A, p. 156
Options\Channel A\Outputs\...\Status OK
Tab. 14.2: Selection of the switching function
property setting description
switching function NO The binary output is energized if the switching condition is met
(invalid measured value) and de-energized if the switching con-
dition is not met (valid measured value).
NC The binary output is de-energized if the switching condition is
met (invalid measured value) and energized if the switching
condition is not met (valid measured value).
Options\Channel A\Outputs\...\Output info