14 Outputs
FLUXUS WD 14.1 Configuration of a binary output
113
UMFLUXUS_WDV1-3EN, 2018-07-16
Select the source item to be transmitted to the output by the channel.
14.1.1 Output of a status value
Switching condition (status OK)
Select the switching function, see Tab. 14.2.
•Press ENTER.
.
If no measurement is carried out, all binary outputs are de-energized, independent of the programmed switching condition.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing key or further information is displayed.
Press ENTER.
If the transmitter possesses a switchable current output, it is displayed whether it is active or passive.
Tab. 14.1: Output of status values or event values
source item status value event value
physical quantities Flow quantities x
Totalizers x
Fluid properties x
Pulse x
Sound speed x
results Event trigger x
Source item
Switching function
(Status OK)
Output info
Test signal
Test mea. range
see annex A, p. 156
Options\Channel A\Outputs\...\Status OK
Tab. 14.2: Selection of the switching function
property setting description
switching function NO The binary output is energized if the switching condition is met
(invalid measured value) and de-energized if the switching con-
dition is not met (valid measured value).
NC The binary output is de-energized if the switching condition is
met (invalid measured value) and energized if the switching
condition is not met (valid measured value).