15  Data logger
15.1  Configuration of the data logger FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
118
Starting time
It is possible to set a starting time if it is necessary to synchronize the storing of measured values for several transmitters.
• Select the menu item Configuration in the program branch Special functions\Data logger.
• Press ENTER until the menu item Start storing is displayed.
• Select the moment at which the storing has to start.
Storage rate
The storage rate indicates the interval how often the measured values are transmitted or stored. If a start time for storing
the measured values is defined, a storage rate has to be entered.
• Select the menu item Configuration in the program branch Special functions\Data logger.
• Press ENTER until the menu item Storage rate is displayed.
• Select a storage rate from the scroll list.
• Press ENTER.
•If User-defined is selected, enter the storage rate.
• Press ENTER.
Special functions\
Data logger\Configuration
\Start storing
display description
Immediately The storing starts immediately.
Full 5 minutes The storing starts in the next full 5 minutes.
Full 10 minutes The storing starts in the next full 10 minutes.
Full 15 minutes The storing starts in the next full 15 minutes.
Full 30 minutes The storing starts in the next full 30 minutes.
Full hour The storing starts in the next full hour.
Event-based The storing starts once a defined event has been occurred.
actual time: 09:06 am
setting: Full 10 minutes
The storing starts at 09:10 am.
Make sure that the time settings of all transmitters are synchronized.
Special functions\
Data logger\Configuration
\Storage rate