13 User modes
FLUXUS WD 13.1 StandardUser mode
105
UMFLUXUS_WDV1-3EN, 2018-07-16
Enter the measured values of the transmitter.
Press ENTER.
Enter the measured values of the reference flowmeter.
Press ENTER.
Repeat the input for all pairs of measured values.
Press ENTER after each input.
•Select Yes to also apply the calibration curve for negative flow velocities. Select No if it is not to be used for negative flow
velocities.
Start in measuring mode
For some application it is necessary to start the measurement in a particular measuring mode.
Select the menu item Special settings in the program branch Options.
Press ENTER until the menu item Start in meas. mode is displayed.
The menu item Start in meas. mode will only be displayed if FastFood or NoiseTrek mode is enabled.
•Select TransitTime, FastFood, NoiseTrek to start the measurement in the corresponding mode.
Press ENTER.
•Select Yes to keep always the same the measuring mode. Select No to select another measuring mode by pressing key
during the measurement.
The measuring modes FastFood and NoiseTrek can only be selected if they are available and activated in the transmitter.
Options\Special settings\Multi-point calibration\Point x=act. value
Options\Special settings\Multi-point calibration\Point x=set value
Options\Special settings\Multi-point calibration\Bidirectional use
Options\Special settings\Start in meas. mode
Options\Special settings\Start in meas. mode\Only ... mode