13 User modes
13.3 SuperUser Mode and SuperUser ext. mode FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
110
Linear calibration
It is possible to define a correction of the flow velocity:
v
cor
= m · v + n
where
All quantities derived from the flow velocity will be calculated with the corrected flow velocity.
Select the menu item Special settings in the program branch Options.
Press ENTER until the menu item Linear calibration is displayed.
•Select Off to measure without linear calibration.
•Select Default if no user-defined inputs are to be made.
•Select On to define the values for the calibration.
Press ENTER.
Enter the factor for the linear calibration.
Press ENTER.
Enter the offset for the linear calibration.
Press ENTER.
v measured flow velocity
m factor, range: -2...+2
n offset, range: -12…+12 cm/s
v
cor
corrected flow velocity
Notice!
It will not be displayed that the correction of the flow velocity is active during the measurement.
Options\Special settings\Linear calibration
Options\...\Factor
Options\...\Offset
Example
factor: 1.1
offset: -10 cm/s = -0.1 m/s
If a flow velocity v = 5 m/s is measured, before the calculation of the derived quantities, it will be corrected as follows:
v
cor
= 1.1 · 5 m/s - 0.1 m/s = 5.4 m/s
Example
factor: -1
offset: 0
Only the sign of the measured values changes.
Notice!
The correction data will only be stored when a measurement is started.
If the transmitter is switched off without starting a measurement, the entered correction data will be lost.