17 Advanced functions
FLUXUS WD 17.3 Calculation channels
131
UMFLUXUS_WDV1-3EN, 2018-07-16
17.3.4 Measurement with calculation channels
Select the program branch Measurement.
Press ENTER.
Activate the necessary channels. The calculation channels are activated or deactivated the same way as measuring
channels.
Press ENTER.
17.3.5 Advanced diagnosis
An advanced diagnosis is available for the calculation channels in the ExperUser, SuperUser and SuperUser ext. mode.
The aim of the advanced diagnosis is to detect errors on individual measuring channels. The values of the advanced diag-
nosis can either be transmitted via the outputs of the transmitter or defined as source of the event trigger.
Selection of the calculation channel
Select the program branch Options.
Press ENTER.
Select the calculation channel (here: Channel Y).
Press ENTER.
Assignment of an output
•Select Outputs.
Press ENTER.
Select the output to be assigned to the calculation channel (here: Current I1(--)).
Press ENTER.
The scroll list contains all available outputs of the transmitter:
Current Ix (--)
Voltage Ux (--)
Binary Bx (--)
Frequency Fx (--)
If the output has already been assigned to a channel, it is displayed as follows: Current I1 (Y).
•Select Yes to change the settings for an already assigned output or to assign a new output.
•Select No to cancel the assignment and to return to the previous menu item.
Press ENTER.
Measurement\Select channels
Notice!
If a measuring channel that is needed for an activated calculation channel is deactivated, no value is output for the
calculation channel.
Options\Channel Y
Options\Channel Y\Outputs\Current I1(--)
Options\Channel Y\Outputs\Current I1\Enable I1