17 Advanced functions
FLUXUS WD 17.7 Event trigger
139
UMFLUXUS_WDV1-3EN, 2018-07-16
Definition of the switching condition
Select the switching condition.
Press ENTER.
Definition of the holding behavior
Select the type of the holding behavior.
Press ENTER.
Definition of trigger limits
It is important to enter the limits, when the event trigger have to toggle.
Enter the upper limit MAX (x>limit).
Press ENTER.
Enter the lower limit MIN (x<limit).
Press ENTER.
It is possible to define a hysteresis to avoid a constant switching of the event trigger.
The event trigger will be activated if the measured values exceed the upper limit. It will be deactivated if the measured val-
ues fall below the lower limit.
Enter the value for the hysteresis.
If zero is entered, no hysteresis is used.
Press ENTER.
Enter the value for the range center in which the event trigger has to be switched.
Press ENTER.
Enter the value for the range width in which the event trigger has to be switched.
Press ENTER.
Options\Event trigger\Enable Rx\Source item\...\Function
Options\Event trigger\Enable Rx\Source item\...\Type
Options\Event trigger\Enable Rx\Source item\...\Trigger value
Options\Event trigger\Enable Rx\Source item\...\Hysteresis
Example
MAX (x>limit): 30 m³/h
Hysteresis: 1 m³/h
The event trigger will switch at a value of 30.5 m³/h and higher. The event trigger will be deactivated at a value of
29.5 m³/h.
Options\Event trigger\Enable Rx\Source item\...\Range center
Options\Event trigger\Enable Rx\Source item\...\Range width
Example
Function: Out of range
Range center: 100 m³/h
Range width: 40 m³/h
The event trigger will switch if the measured value is below 80 m³/h or above 120 m³/h.