3 General principles
3.4 Undisturbed flow profile FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
18
Tab. 3.1: Recommended distance from disturbance sources
D – nominal pipe diameter at the measuring point
l – recommended distance between disturbance source and transducer position
disturbance source: 90° elbow
inlet: l 10 D outlet: l 5D
disturbance source: 2 × 90° elbows on different planes
inlet: l 40 D outlet: l 5D
disturbance source: valve
inlet: l 40 D outlet: l 5D
disturbance source: reducer
inlet: l 10 D outlet: l 5D
disturbance source: pump
inlet: l 20 D