7 Connection
7.1 Transducers FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
60
7.1.5 Transducer module (SENSPROM)
The transducer module contains important transducer data for the operation of the transmitter with transducers.
If transducers are replaced or added, the sensor module must also be replaced or added.
The sensor module is connected to the terminals of the transmitter, see Fig. 7.7.
Disconnect the transmitter from the power supply.
Connect the transducer module to the corresponding terminal, see Fig. 7.7. Terminal X2 is assigned to the transducers
of measuring channel A, terminal X3 to the transducers of measuring channel B.
Connect the transmitter to the power supply.
Enter all parameters of the program branch Parameters.
Start the measurement.
Notice!
The serial numbers of the transducer module and the transducer must be identical. A wrong or incorrectly connected
sensor module will lead to incorrect measured values or to a measurement failure.
Fig. 7.7: Sensor module
1 transducer module (measuring channel A)
2 transducer module (measuring channel B)
A
B
SNAP
DISP
DISP
MODEFAST
MUX
NEXT
Q
ON
Q+Q-
3x Q
OFF
LAN
USB
A+
B-
P1+
P2+
P3+
P4+
P5a
P6a
P7a
S
S
P1-
P2-
P3-
P4-
P5b
P6b
P7b
X2 X3
X_AV
KL11
KL12
KL7KL8
KL9KL10
KL4
X_AR X_BV X_BR
T1A
T1B
S1
T2A
T2B
T3A
T3B
S3
T4B
T4A
T1a
T1b
S1
T2a
T2b
T3a
T3b
S3
T4b
T4a
N(-)
PE
L(+)
AV
AVS
AGN
ARS
AR
BV
BVS
BGN
BRS
BR
1 2