7 Connection
7.2 Power supply FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
62
7.2.1 Cable connection
Transmitter with stainless steel housing
Remove the blind plug to connect the cable to the transmitter.
Prepare the cable with a cable gland. The used cable has to have a wire cross-section of 0.25…2.5 mm². The outer
diameter of the cable may not exceed 7.6 mm.
Push the cable through cap nut, compression part, basic part and sealing ring (sealing ring: only for cable gland M20, not
for cable gland 1/2 NPS).
Insert the cable into the housing of the transmitter.
Fix the cable gland by screwing the cap nut onto the basic part.
Fix the transducer cable by tightening the cable gland with the ferrite nut.
Connect the cable to the terminals of the transmitter.
Transmitter with aluminum housing
Remove the blind plug to connect the cable to the transmitter.
Prepare the cable with a cable gland.
The used cable has to have a wire cross-section of 0.25…2.5 mm².
Push the cable through the cap nut, compression part and basic part of the cable gland.
Insert the cable into the housing of the transmitter.
Screw the sealing ring side of the basic part into the transmitter housing.
Fix the cable gland by screwing the cap nut onto the basic part.
Connect the cable to the terminals of the transmitter.
Tab. 7.6: Terminal assignment
terminal connection AC connection DC
PE earth earth
N(-) neutral -
L(+) phase 100...230 V AC, 50...60 Hz +
fuse 1 A, time-lag 1.6 A, time-lag
Fig. 7.10: Cable gland for stainless steel housing
1 cap nut
2 compression part
3 basic part
4 sealing ring: only for cable gland M20, not for cable gland 1/2 NPS
5 housing wall
6 ferrite nut