7 Connection
7.3 Outputs FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
64
switchable current output
The switchable current output
is switched to active or passive
in the menu item Special
functions\Outputs.
active current output
R
ext
< 350
passive current output
U
ext
=
8...30 V
U
ext
> 0.021 A
.
R
ext
[] + 8 V
example:
U
ext
=12V
R
ext
190
binary output (optorelay) circuit 1 U
ext
26 V
I
c
100 mA
circuit 2
RS485 120
termination resistor
Tab. 7.7: Circuit of the outputs
output transmitter external circuit remark
internal circuit connection
The number, type and the connections of the outputs depend on the order.
R
ext
is the sum of all ohmic resistances in the circuit (e.g., resistance of the conductors, resistance of the amperemeter/voltmeter).