9 Measurement
FLUXUS WD 9.2 Measurement settings
83
UMFLUXUS_WDV1-3EN, 2018-07-16
Measuring range
The sign of the measured value and the measuring range are determined.
•Select Sign if the sign of the measured values is to be considered for the output.
•Select Absolute value if the sign of the measured values is not to be considered for the output.
Enter the lowest expected measured value. The unit of measurement of the source item will be displayed.
Start of meas. range is the value assigned to the value Output MIN of the output range.
Enter the highest expected measured value. The unit of measurement of the source item will be displayed.
End of meas. range is the value assigned to the value Output MAX of the output range.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing key or further information is displayed.
Press ENTER.
If the transmitter possesses a switchable current output, it is displayed whether it is active or passive.
Output function test
The function of the output can now be tested.
Connect an external measuring instrument to the terminals of the output.
•Select Yes to test the output. Select No to display the next menu item.
Press ENTER.
Enter a test value. It has to be within the output range.
Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
•Select Repeat to repeat the test, Finish to display the next menu item.
Press ENTER.
•Select Yes to test the assignment of the measured value to the output signal. Select No to display the next menu item.
Press ENTER.
Options\Channel A\Outputs\...\Measured values\Absolute value
Options\Channel A\Outputs\...\Start of meas. range
Options\Channel A\Outputs\...\End of meas. range
Options\Channel A\Outputs\...\Output info
Options\Channel A\Outputs\...\Test signal
Options\Channel A\Outputs\...\Enter test value
Options\Channel A\Outputs\...\Test mea. range