5730A
Operators Manual
6-4
Commands
Table 6-4 summarizes the commands by function. Table 6-5 provides protocol details of
the remote commands. Table 6-6 provides protocol details of the three special commands
available for serial remote control only. The commands duplicate almost all activities that
can be initiated from the front panel in local operation. Separate headings for each
command in the tables provide the parameters and responses (if any), and an example for
cases in which the parameters are not self-explanatory.
Multiple Commands
Controllers may send commands all at once, or one at a time. To set the output to 100 mV
dc and then place the 5730A Calibrator into operate, two separate commands may be
used:
OUT 100 MV <CR/LF>
OPER <CR/LF>
Or enter it as a compound command on a single line:
OUT 100 MV; OPER <CR/LF>
Coupled Commands
When a compound command follows another command, there is a possibility that a fault
can occur because of the action of the first command. Such commands are called coupled
commands. An example of coupled commands that cause a fault is:
OUT 1V, 1 MHZ
followed by the compound command:
OUT 100V” ; OUT 100 HZ.”
The second command causes a fault because when the 5730A Calibrator encounters
OUT 100V”, it is interpreted as 100V, 1 MHz, which is out of range of the Calibrator.
If only the second compound command was sent, there would have been no fault and the
Calibrator would be set to 100 V at 100 Hz.
Another example is:
OUT 1V ; RANGELCK ON
followed by:
OUT 10V ; RANGELCK OFF.”
In this case, when the Calibrator interprets the first half of the second command, it causes
an error since the previous command locked the range on 2.2 V.
Any command that affects the state of the output is a coupled command, for example
order-dependent. These commands include all the following: *RST, BOOST,
BTYPE, CAL_CHK, CAL_REF, CAL_STORE, CAL_ZERO, CUR_POST,
DIAG, EXTGUARD, EXTSENSE, FORMAT, INCR, LIMIT, MULT,
NEWREF, OFFSET, OPER, OUT, PHASE, PHASELCK, PHASESFT,
RANGELCK, RCOMP, REMOTE, LOCAL, LOCKOUT, SCALE, STBY,
WBAND, CAL_RNG. Make sure to use appropriate sequences when these commands are
used.

Table of Contents

Other manuals for Fluke 5730A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5730A and is the answer not in the manual?

Fluke 5730A Specifications

General IconGeneral
TypeMultifunction Calibrator
Resistance Range0 to 1 GΩ
AC Voltage Frequency10 Hz to 1 MHz
DisplayColor LCD
InterfacesRS-232, Ethernet
Weight18 kg
DC Voltage Range0 to ±1100 V
Dimensions172 mm H x 432 mm W x 457 mm D (6.8 in H x 17 in W x 18 in D)

Related product manuals