EasyManua.ls Logo

Fluke 5730A - Offset Error; Scale Error

Fluke 5730A
248 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5730A
Operators Manual
4-36
Offset Error
Offset error can be measured directly by finding the Calibrator output that causes a meter
reading of 0 V. This error is called an offset because it reflects a fixed error that is present
in all meter output readings. For example, if a meter reads 0 V when -1.3 mV is applied,
the meter has an offset error of +1.3 mV. Figure 4-7 illustrates this example.
Calibrator
Setting
0
-1.3 mV
Meter Reading
Offset
Error =
+1.3 mV
hhp052.eps
Figure 4-7. Offset Error
Scale Error
Scale error, sometimes referred to as gain error, occurs when the slope of the meter
response curve deviates from one. A meter with only scale error (no offset or linearity
error), will read 0 V when 0 V is applied, but will read a value other than 10 V when
10 V is applied. For example, if a meter reads 19.900 V when 19.903 V is applied, the
meter has a scale error of -3 mV. To isolate scale error, offset error must first be
subtracted. Scale error is then simply the error measured near the full-scale endpoint, or:
Scale Error=
(Reference full scale) - (Adjusted Calibrator Setting)
(Reference full Scale)
Where “Adjusted calibrator setting” is the adjusted output (using the knob) that results in
the UUT correctly reading “nominal full-scale.” Select a point just down from the UUT
full-scale endpoint for nominal full-scale. This keeps the UUT from ranging while
adjustments are made. For example, use 19.9 V as the nominal full-scale for a UUT that
ranges at 20 V.

Table of Contents

Other manuals for Fluke 5730A

Related product manuals