5730A
Operators Manual
1-20
Resistance Specifications
5730A Resistance Specifications
Nominal Value
(Ω)
Absolute Specification of Characterized Value
±5 °C from calibration temperature
[1]
Relative ±1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
±ppm
99 % Confidence Level
0
1
1.9
10
19
100
190
1 k
1.9 k
10 k
19 k
100 k
190 k
1 M
1.9 M
10 M
19 M
100 M
50 μΩ
85
85
23
23
10
10
7
7
6
6
7
8
13
15
33
43
100
50 μΩ
95
95
25
25
11
11
7.2
7.2
7
7
8
10
14
17
37
47
110
50 μΩ
100
100
26
26
11.5
11.5
7.5
7.5
7.5
7.5
9
11
14.5
19
40
50
115
50 μΩ
110
110
27
27
12
12
8
8
8
8
10
12
15
21
46
55
120
50 μΩ
32
25
5
4
2
2
2
2
2
2
2
2
2.5
3
10
20
50
50 μΩ
40
33
8
7
4
4
3
3
3
3
3
3
5
6
14
24
60
95 % Confidence Level
0
1
1.9
10
19
100
190
1 k
1.9 k
10 k
19 k
100 k
190 k
1 M
1.9 M
10 M
19 M
100 M
40 μΩ
70
70
20
20
8
8
5.5
5.5
5
5
5.5
6
10
12
27
35
85
40 μΩ
80
80
21
21
9
9
5.7
5.7
5.5
5.5
7.5
7
11
13.5
31
39
95
40 μΩ
85
85
22
22
9.5
9.5
6
6
6
6
8
8
12
15
34
42
100
40 μΩ
95
95
23
23
10
10
6.5
6.5
6.5
6.5
8.5
8.5
13
18
40
47
100
40 μΩ
27
20
4
3.5
1.6
1.6
1.6
1.6
1.6
1.6
1.6
1.6
2
2.5
8
16
40
40 μΩ
35
26
7
6
3.5
3.5
2.5
2.5
2.5
2.5
2.5
2.5
4
4
12
20
50
Note:
1. Specifications apply to displayed value. 4-wire connections, except 100 MΩ.