5730A
Operators Manual
6-28
Table 6-5. Commands (cont.)
EMULATE
Description
Partially emulates a 5700A or 5720A over the remote interface. This changes the model
number in the *IDN? response. It also accepts BTYPE VB5205 and implements it as if
BTYPE VBNONE was sent, as a Fluke 5205A cannot be connected to the 5730A
Calibrator.
Parameter
(Integer) 5700 selects 5700A emulation; 5720 selects 5720A emulation; anything else
selects normal 5730A behavior.
Example
EMULATE 5720
Sets 5720A emulation.
EMULATE?
Description Returns the state of emulation as set by the EMULATE command.
Parameter None
Response (Integer) 5700 for 5700A emulation, 5720 for 5720A emulation, 0 for normal behavior.
ENETPORT
Description Sequential command. Ignored if not in remote. Sets the Ethernet port number.
Parameter Port number
Example
ENETPORT 3490
Sets the Ethernet port number to 3490.
ENETPORT?
Description Sequential command. Returns the Ethernet port number.
Parameter None
Response Integer
Example
ENETPORT?
Returns 3490 if the Ethernet port number is set to 3490
EOL
Description
Sequential command. Ignored if not in remote. Sets the end of line terminator for
outgoing data for a specified remote port.
Parameter
1. SERIAL, USB, ENET
2. CRLF, CR, LF
Example
EOL ENET, CR
Sets the end of line terminator for Ethernet communication to CR.