Remote Commands and Syntax
Commands 6
6-31
Table 6-5. Commands (cont.)
FORMAT
Description
Sequential command, ignored if not in remote.
Use with extreme care. Restores the contents of the non-volatile memory to factory
defaults. The non-volatile memory holds calibration constants and setup parameters. All
calibration data is lost permanently. The Calibrator security state must be set to OFF: or
an execution fault occurs.
Parameter
ALL (Replaces the whole contents with defaults)
B5725 (Replaces 5725A calibration constants with defaults)
CAL (Replaces all calibration constants with defaults)
RNG (Replaces range calibration constants with defaults)
SETUP (Replaces setup parameters with defaults)
GWADDR
Description
Sequential command. Ignored if not in remote. Sets Ethernet gateway address for LAN
communication when NOT in DHCP mode.
Parameter
Gateway address (quoted string consisting of 4 decimal values bound between 0-255
separated by periods).
Example
GWADDR "129.196.136.1"
Sets the Etherent gateway address to 129.196.136.1
GWADDR?
Description
Sequential command. Returns Ethernet gateway address for LAN communication when
NOT in DHCP mode. When in DHCP mode, the response will be default.
Parameter None
Response String
Example
GWADDR?
Returns 129.196.136.1 if the gateway address is set to 129.196.136.1 and DHCP is not
enabled. Returns default if DHCP is enabled.
ID52120? (returns the number of 52120s connected and their serial numbers)
Description
Sequential command. Returns the number of 52120s connected and their serial
numbers.
Parameter None
Response One or more integers
Example
ID52120?
Returns 0 if no 52120As are connected. Returns 2, 7346432, 8883213 if two 52120A's
are connected with the indicated serial numbers.