6-38
Table 6-5. Commands (cont.)
PHASE?
Description
Sequential command.
Returns the phase angle of the variable phase output signal, with respect to the main
output signal.
Parameter None
Response (Integer) Phase in degrees (0 to 180,-179 to 0, 0 if the output is not ac)
PHASELCK
Description
Overlapped command, ignored if not in remote.
Enables or disables external phase locking, when the Calibrator is supplying an ac
voltage.
Parameter
ON (Activates phase locking)
OFF (Deactivates phase locking)
PHASESFT
Description
Overlapped command, ignored if not in remote.
Enables or disables the variable phase output, when the Calibrator is supplying an ac
voltage.
Parameter
ON (Activates variable phase output)
OFF (Deactivates variable phase output)
POST_52120
Description
Sequential command. Ignored if not in remote. Sets the output terminals for all attached
52120As.
Parameter
1. LO52120, HI52120
Example
POST_52120 LO52120
Selects the low current output terminals for all attached 52120As.
POST_52120?
Description
Sequential command. Returns the selected output terminals for all attached 52120As.
Parameter
None
Response
String
Example
POST_52120?
Returns HI52120 if the high current output terminals are selected for all attached
52120As.