5730A
Operators Manual
6-40
Table 6-5. Commands (cont.)
RCOMP
Description
Overlapped command, ignored if not in remote.
While a resistance output 19 kΩ or lower is selected, RCOMP activates or deactivates
two-wire ohms compensation circuitry.
Parameter
ON (Turns on the two-wire compensation circuitry)
OFF (Turns off the two-wire compensation circuitry)
REFOUT?
Description
Sequential command.
Returns the value of the reference, which is the output value of the Calibrator the last
time a new reference was established with an OUT, NEW_REF, or MULT.
Parameter None
Response
1. (Float) Output amplitude
2. (String) Units (V, A, OHM, or DBM)
3. (String) Frequency (0 if dc)
REM_MODE
Description
Sequential command. Ignored if not in remote. Sets the response type for a specified
remote port.
Parameter
1. SERIAL, USB, ENET
2. COMP,TERM
Example
REM_MODE SERIAL, COMP
Sets the response type for serial communication to COMPUTER
REM_MODE ENET, TERM
Sets the response type for Ethernet communication to TERMINAL.
REM_MODE?
Description Sequential command. Returns the response type for a specified remote port.
Parameter 1. SERIAL, USB, ENET
Response String
Example
REM_MODE? SERIAL
Returns TERM if the serial response type has been set to TERMINAL mode
RPT_STR
Description
Sequential command.
Loads the user report string. The user report string can be read on the Display in local
operation, and appears on calibration reports. The Calibrator secure state must be set to
off.
Parameter String of up to 40 characters