Operators Manual
7-14
1. Touch Setup Menu>Calibration>Range Adjustment to bring up the Range
Adjustment menu shown below:
Range
Adjustment
Exit
DCI
ACV
ACI
DCV
Base 19 Values
Resistance
Base 10 Values
Setup Menu Calibration
hhp189.eps
2. Touch DCV to show the next menu shown below:
Range
Adjustment
Setup Menu Calibration
220 mV 1100 V
2.2 V
11 V
DC V Range Calibration
Cancel
22 V
220 V
hhp190.eps