5730A
Operators Manual
1-18
1000
100
10
0
3.5 V
10000
50 Hz10 Hz 1 kHz 100 kHz 1 MHz 30 MHz
Voltage
Frequency
40 Hz 30 Hz
2.2 x 10
7
V-Hz
15 Hz
Volt-Hertz Capability
Figure A.
5725A Amplifier:
Range
Frequency
(Hz)
Stability
±1 °C
[1]
24 Hours
Temperature Coefficient
Adder
Load Regulation
[2]
Distortion
Bandwidth
10 Hz -10 MHz
±(% output)
10 - 40 °C
0 - 10 °C and
40 - 50 °C
(ppm output + mV) ±(ppm output) / °C ±(ppm output + mV)
150 pF 1000 pF
1100 V
40 - 1 k
1 k - 20 k
20 k - 50 k
50 k - 100 k
10 + .5
15 + 2
40 + 2
130 + 2
5
5
10
30
5
5
10
30
10 + 1
90 + 6
275 + 11
500 + 30
0.10
0.10
0.30
0.40
0.10
0.15
0.30
0.40
Notes:
1. Stability specifications are included in Absolute specification values for the primary specifications.
2. The 5725A will drive up to 1000 pF of load capacitance. Absolute specifications include loads to 300 pF and 150 pF as shown
under "Load Limits." For capacitances up to the maximum of 1000 pF, add "Load Regulation."
Voltage Range Maximum Current Limits Load Limits
2.2 V
[2]
22 V
220 V
50 mA, 0 °C-40 °C
20 mA, 40 °C-50 °C
>50 Ω,
1000 pF
1100 V 6 mA 600 pF
5725A Amplifier:
40 Hz-5 kHz 50 mA
1000 pF
[1]
1100 V 5 kHz-30 kHz 70 mA 300 pF
30 kHz-100 kHz 70 mA
[3]
150 pF
Notes:
1. The 5725A will drive up to 1000 pF of load capacitance. Absolute specifications include loads to 300 pF and 150 pF as shown
under "Load Limits." For capacitances up to the maximum of 1000 pF, add "Load Regulation."
2. 2.2 V Range, 100 kHz-1.2 MHz only: Absolute specifications cover loads to 10 mA or 1000 pF. For higher loads, load regulation
is added.
3. Applies from 0 °C to 40 °C.
Output Display Formats ....................................... Voltage or dBm, dBm reference 600 Ω.
Minimum Output ................................................... 10 % on each range
External Sense ...................................................... Applicable for 2.2 V, 22 V, 220 V, and 1100 V ranges; 5730A
<100 kHz, 5725A <30 kHz. Specifications are the same as internal
sense.