5730A
Operators Manual
1-26
AC Current Secondary Performance Specifications and Operating Characteristics
Range
Frequency
(Hz)
Stability ±1 °C
[1]
24 Hours
Temperature
Coefficient
[2]
Compliance
Limits
(V rms)
Maximum
Resistive
Load
For
Published
Specification
[3]
(Ω)
Noise and
Distortion
(Bandwidth
10 Hz - 50 kHz
<0.5V Burden)
10 - 40 °C
0 - 10 °C and
40 - 50 °C
±(ppm output +
nA)
±(ppm output + nA)/ °C ±(% output + μA)
220 μA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
150 + 5
80 + 5
30 + 3
50 + 20
400 + 100
50 + 5
20 + 5
4 + 0.5
10 + 1
20 + 100
50 + 5
20 + 5
10 + 0.5
20 + 1
20 + 100
7 2 k
0.05 + 0.1
0.05 + 0.1
0.05 + 0.1
0.25 + 0.5
00.5 + 1
2.2 mA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
150 + 5
80 + 5
30 + 3
50 + 20
400 + 100
50 + 5
20 + 4
4 + 1
10 + 100
50 + 400
50 + 5
20 + 4
10 + 2
20 + 100
50 + 400
7 800
0.05 + 0.1
0.05 + 0.1
0.05 + 0.1
0.25 + 0.5
00.5 + 1
22 mA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
150 + 50
80 + 50
30 + 30
50 + 500
400 + 1000
50 + 10
20 + 10
4 + 10
10 + 500
50 + 1000
50 + 10
20 + 10
10 + 20
20 + 400
50 + 1000
7 80
0.05 + 0.1
0.05 + 0.1
0.05 + 0.1
0.25 + 0.5
00.5 + 1
Hz
±(ppm output +
μA)
±(ppm output + μA)/°C
220 mA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
150 + 0.5
80 + 0.5
30 + 0.3
50 + 3
400 + 5
50 + 0.05
20 + 0.05
4 + 0.1
10 + 2
50 + 5
50 + 0.05
20 + 0.05
10 + 0.1
20 + 2
50 + 5
7 8
0.05 + 10
0.05 + 10
0.05 + 10
0.25 + 50
00.5 + 100
2.2 A
20 - 1 k
1 k - 5 k
5 k - 10 k
50 + 5
80 + 20
800 + 50
4 + 1
10 + 5
50 + 10
10 + 1
20 + 5
50 + 10
1.4
[4]
0.8
0.5 + 100
0.3 + 500
1 + 1 mA
5725A Amplifier:
±(% output)
11 A
40 - 1 k
1 k - 5 k
5 k - 10 k
75 + 100
100 + 150
200 + 300
20 + 75
40 + 75
100 + 75
30 + 75
50 + 75
100 + 75
3 3
0.05
[5]
0.12
[5]
0.5
[5]
Notes:
Maximum output from 5730A terminals is 2.2 A. Specifications for 220 μA and 2.2 mA ranges are increased by a factor of 1.3, plus
2 μA when supplied through 5725A terminals. Specifications are otherwise identical for all output locations.
1. Stability specifications are included in the Absolute values for the primary specifications.
2. Temperature coefficient is an adder to specifications that does not apply unless operating more than ±5 °C from calibration
temperature.
3. For larger resistive loads multiply accuracy specifications by: (
ionspecificat publishedfor load maximum
load actual
)
2
4. 1.5 V compliance limit above 1 A. 5725A Amplifier may be used in range-lock mode down to 1 A.
5. For resistive loads within rated compliance voltage limits.