In dc current output, the selections available at the bottom are:
Current Output
External Guard
Boost
Wideband
Setup Menu
In addition, the Offset and Scale functions are available. The output range can be locked
in dc current.
Current Output selects one of three locations for non-boosted (current sourced by the
Calibrator, for example) current output:
OUTPUT binding post (NORMAL, which is the OUTPUT HI binding post, also the
default setting)
AUX, which is the AUX CURRENT OUTPUT binding post
5725A which is the 5725A Amplifier binding posts. (The 5725A must be turned on,
but not necessarily activated.)
Non-boosted current from the Calibrator is not available at the 52120A amplifier binding
posts.