5730A
Operators Manual
4-24
If the “Ω Value Table” selection is used, touch the desired resistance value and to
select it. To exit the menu, touch the “x” on the top right of the table.
5. Push or if necessary.
6. Push .
7. Push . If a resistance value that is unavailable (as in the subsequent example
of 490 Ω) is entered, the display prompts to try again.
+/- 0.00050mV (1 year) 213 days since calibration
12/21/12 10:52am
External
Sense
OFF
External
Guard
OFF
Wideband
OFF
Setup
Menu
Standby
mV
0.00000
Auto Range 220 mV
External
Sense
OFF
Error 820: Calibrator Cannot Source
That Value.
OK
hhp123.eps
8. After a valid resistance is specified and is pushed, the Calibrator clears the
entry from the entry bar near the bottom and shows in the larger area above.
9. Push . This resistance is now available at the output terminals.
dBm
kHz
1.0000
42.00000
+/- 10.0 ppm (1 year) 213 days since calibration
3/15/2013 10:52am
External
Sense
OFF
External
Guard
OFF
2 Wire
Comp
ON
Wideband
OFF
Setup
Menu
Standby
Expected Value 1.9 kΩ
1.9000096
External
Sense
OFF
hhp124.eps