9190A
Operators Manual
1-6
Calibration and Repair Information
To schedule and send the Calibrator to Fluke for calibration or repair:
1. Contact the Fluke Calibration Service Center in your area to schedule the calibration
or repair (see “Contact Fluke Calibration” on page 1-5).
2. Pack and secure the Calibrator in a shipment box with a minimum of 2 inches of
packing around the Calibrator to prevent damage.
3. Send the Calibrator to the Service Center.
Specifications
Base Unit Specifications
Temperature Range at 23 °C ................................ –95 °C to 140 °C (–139 °F to 284 °F)
Display Accuracy .................................................. ±0.2 °C Full Range
Accuracy with External Reference
[3]
................. ±0.05 °C Full Range
Stability .................................................................. ±0.015 °C Full Range
Axial Uniformity at 40 mm (1.6 in) ....................... ±0.05 °C Full Range
Radial Gradient ..................................................... ±0.01 °C Full Range
Loading Effect
(with a 6.35 mm reference
probe and three 6.35 mm probes) ...................... ±0.006 °C Full Range
(versus display with
6.35 mm probes)................................................. ±0.25 °C at –95 °C
±0.10 °C at 140 °C
Operating Conditions ........................................... 0 °C to 35 °C, 0 % to 90 %
RH (non-condensing) < 2000 m altitude
Environmental conditions for all
specifications except temperature range ........... 13 °C to 33 °C
Immersion (Well) Depth ........................................ 160 mm (6.3 in)
Well Diameter ........................................................ 30 mm (1.18 in)
Heating Time
[1]
..................................................... –95 °C to 140 °C: 40 min
Cooling Time
[1]
..................................................... 23 °C to –90 °C: 80 min
23 °C to –95 °C: 90 min
140 °C to 23 °C: 60 min
Stabilization Time
[2]
............................................. 15 min
Resolution ............................................................. 0.01 °
Display ................................................................... LCD, °C or °F user selectable
Size (H x W x D) ..................................................... 480 mm x 205 mm x 380 mm (18.8 in x 8.0 in x 14.9 in)
Weight .................................................................... 16 kg (35 lb)
Power Requirements ............................................ 100 V to 115 V (±10 %) 50/60 Hz, 575 W
200 V to 230 V (±10 %) 50/60 Hz, 575 W
System Fuse Ratings............................................ 115 V: 6.3 A T 250 V
230 V: 3.15 A T 250 V
4–20 mA Fuse (-P model only) ............................. 50 mA F 250 V
Computer Interface ............................................... RS-232, USB Serial, and 9930 Interface-it Temperature Calibration
Software included
Safety ..................................................................... IEC 61010-1, Installation Category II, Pollution degree 2
Electromagnetic Environment ............................. IEC 61326-1: Basic
Refrigerants
R32 (Difluoromethane) ....................................... < 20 g, ASHRAE Safety Group A2L
R704 (Helium) ..................................................... < 20 g, ASHRAE Safety Group A1