Operation
Operation 2
2-33
Operation
Insert Installation and Removal Procedure
Warning
For safe operation and maintenance of the product, do not
remove inserts when the block temperature indicator is
illuminated.
To install the insert:
1. Remove the rubber well insulation cap.
2. Stand the insert upright.
3. Lock the insert tool into the attachment holes.
4. Pick up the insert and slowly and carefully lower the insert into the well.
To remove the insert:
1. Let the Calibrator cool to below 50 °C and make sure the high temperature indicator
is not illuminated.
2. Use the insert tool to lift the insert from the well. Put the insert on a metal or ceramic
fiber surface or container made to set hot materials on.
Note
If the insert has not been cleaned periodically (see “Clean the Insert” in
Chapter 4), hard-water deposits can form on the insert and make it difficult
to remove. If the insert cannot be removed, try to remove the insert as
follows:
a. Put the Calibrator in a cold environment that is less than 21 °C.
b. Use the procedure in “Set the Temperature SETPOINT Manually” on page 2-36
to set the temperature to 100 °C.
c. When the temperature is between 50 °C and 70 °C, use the insert tool to lift the
insert from the well. Put the insert on a metal or ceramic fiber surface or
container made to set hot materials on.
d. If the insert cannot be removed, turn off the Calibrator and contact Fluke
Calibration (see “Contact Fluke Calibration” in Chapter 1).
3. Put the rubber well insulation cap back on the well.