Operation
Menus and Menu Navigation 2
2-15
Program Setup
Table 2-12 shows and describes the menu selections on the Program Setup menu.
Table 2-12. Program Setup Menu
66.03°C
PROGRAM SETUP
TEST STATUS: OFF
RECORD DATA: NO
TEST ID: 0
PROG
MENU
RUN
PROG
MENU
1
2
3
gzs011.eps
Item Function
TEST STATUS starts (ON) or stops (OFF) a Program.
RECORD DATA
[-P Only]
turns on (YES) or turns off (NO) the data record function of the Program.
TEST ID
[-P Only]
is a unique identification of the Program. The identification can be an alpha-
numeric entry up to 16 characters in length.
Ramp/Soak
Table 2-13 shows and describes the menu selections on the Ramp/Soak Setup menu.
Table 2-13. Ramp/Soak Setup Menu
66.03°C
RAMP/SOAK SETUP
NO. SETPOINTS: 8
SOAK TIME: 15 MIN
NO. CYCLES: 1
DIRECTION:
PASS TOLERENCE: 1.80 °C
SET
POINTS
PROG
MENU
RAMP/
SOAK
MENU
1
2
3
4
5
6
gzs013.eps
Item Function
NO. SETPOINTS is the number of SETPOINTS for a Program. Eight SETPOINTs can be set for
the program. Set the maximum number of SETPOINTS necessary.
SOAK TIME is the duration (in minutes) that each of the programmed SETPOINTS temperature is
maintained. The time starts when the temperature settles to within the specified stability. The
Stability Limit is set in the Temperature Setup menu (see page 2-12).
NO. CYCLES is the number of times that the Program is repeated.