9190A
Operators Manual
2-36
Set Temperature
A set temperature value is referred to as a “SETPOINT”. The Calibrator uses the
SETPOINT value to know what temperature to heat or cool to. Use the procedures in the
subsequent sections to manually set a SETPOINT, load a SETPOINT preset, or change a
SETPOINT preset.
Set the Temperature SETPOINT Manually
To adjust the Calibrator temperature SETPOINT manually (see Figure 2-10):
1. Push .
2. Push .
3. Push and to highlight a digit then push to increase the digit or push to
decrease the digit.
4. Push twice to save the SETPOINT temperature and start the Calibrator.
66.03°C
SET TEMPERATURE SETPOINT
SELECT
PRESET
SETPOINT:
+080.00 °C
SAVE/
DISABLE
gzs035.eps
Figure 2-10. Set Temperature SETPOINT
Set the Temperature with a Preset SETPOINT
The Calibrator stores eight user-configured preset SETPOINT temperatures to help
quickly set a temperature. To change a SETPOINT preset see “Change a SETPOINT
Preset” on page 2-37.
To load and use a preset SETPOINT (see Figure 2-11):
1. Push on the Main screen.
2. Push .
3. Push or to move between presets (1 through 8).
4. Push to select a preset.
66.03°C
SELECT PRESET
EDIT
PRESET
PRESET:
(SETPOINT: 80.00 °C)
gzs034.eps
Figure 2-11. Preset Selection