Table 2-13. Ramp/Soak Setup Menu (cont.)
Item Function
DIRECTION controls whether the SETPOINTS are sequenced in one direction “→” (ascending
from 1 to 8) or in both directions “ ” (ascending from 1 to 8 and then descending from 8 to 1)
before the sequence is repeated. If the both directions option is selected, the Program sequences
from the first SETPOINT to the last and then reverses direction sequencing from the last to the
first.
PASS TOLERANCE
[-P Only]
is the allowable tolerance condition for the test and is used to highlight
test points that have large errors.
SETPOINTS MENU opens a menu to set each of the SETPOINTS for the Program. Only the
number of SETPOINTS defined by NO. SETPOINTS will be displayed.
Test Result (-P Only)
Table 2-14 shows and describes the menu selections on the Test Result menu. This menu
is available on the -P Model only.
Table 2-14. Test Result Menu
66.03°C
TEST RESULT MENU
VIEW
TEST
PRINT
TEST
ERASE
TEST
PROG
MENU
TEST
RESULT
MENU
1
2 3
gzs012.eps
Item Function
VIEW TEST
[-P Only]
opens a menu where test results can be viewed.
PRINT TEST
[-P Only]
opens a menu to select a programs results and transmit them to a PC to be
printed. The data is transmitted from the RS-232 or USB port in ASCII format. Terminal emulator
or other software can be used to receive the data and save it to a text file on a computer.
ERASE TESTS
[-P Only]
opens a menu to erase test results.