Operation
Menus and Menu Navigation 2
2-25
TC Setup
Table 2-26 shows and describes the menu selections on the TC Setup menu.
Table 2-26. TC Setup Menu
66.03°C
TC SETUP
INPUT
SETUP
SETUP
INPUT
TC
SETUP
TC TYPE: TC-E
MENU
1
gzs028.eps
Item Function
TC TYPE sets the thermocouple conversion type. Standard thermocouple conversions include
types C, E, J, K, L, M, N, R, S, T, and U.
Notes
When mV is selected, the TC Type shows the measurement in uncompensated
millivolts rather than temperature.
The cold junction compensation is measured at the input, automatically computed,
then read by the Calibrator.
mA Setup
Table 2-27 shows and describes the menu selections on the mA Setup menu.
Table 2-27. mA Setup Menu
66.03°C
mA SETUP
INPUT
SETUP
SETUP
INPUT
mA
SETUP
LOOP POWER: ENABLE 24V
MENU
1
gzs029.eps
Item Function
LOOP POWER turns on or turns off the 24 V loop power.