Reference Input Calibration (-P Only)
Table 2-20 shows and describes the menu selections on the Reference Input Calibration
menu. This menu is available on the -P Model only.
Table 2-20. Reference Input Calibration Menu
66.03°C
REFERENCE INPUT CALIBRATION
REF CO: -0.0001
CAL
REF
SYSTEM
MENU CALIB
REF C100: 0.0028
INPUT CAL DATE: 2012-05-01
MENU
1
2
3
gzs021.eps
Item Function
REF C0
[-P Only]
adjusts the first calibration point for the reference resistance.
REF C100
[-P Only]
adjusts the second calibration point for the reference resistance.
INPUT CAL DATE
[-P Only]
is the calibration date for the readout.
TC Input Calibration (-P Only)
Table 2-21 shows and describes the menu selections on the TC Input Calibration menu.
This menu is available on the -P Model only.
Table 2-21. TC Input Calibration Menu
66.03°C
TC INPUT CALIBRATION
TC CO: -0.0386
CAL
UUT
SYSTEM
MENU CALIB
TC C100: -0.0992
TC CRJ: 3.7754
CAL TC
INPUT CAL DATE: 2012-05-01
MENU
1
2
3
4
gzs022.eps
Item Function
TC C0
[-P Only]
adjusts the measurement at 0 mV.
TC C100
[-P Only]
adjusts the measurement at 100 mV.
TC CRJ
[-P Only]
adjusts the reference junction temperature measurement.
INPUT CAL DATE
[-P Only]
is the calibration date for the TC input.