Change a Preset SETPOINT
To change a preset SETPOINT (see Figure 2-12):
1. Push on the Main screen.
2. Push .
3. Push .
4. Push or to move between presets (1 through 8). Highlight the preset to change
then push .
5. Push and to highlight a digit then push to increase the digit or push to
decrease the digit.
6. Push to save the preset.
66.03°C
EDIT PRESETS
PRESET 1:
PRESET 2:
PRESET 3:
PRESET 4:
+0080.00 °C
+45.00 °C
+26.00 °C
-26.00 °C
gzs036.eps
Figure 2-12. Edit SETPOINT Preset
Cancel or Stop Temperature Change
To cancel or stop temperature change:
1. Push on the Main screen.
2. Push .
Set a Soft Cutout
Soft Cutout is an upper-heat limit that protects the probes. The Hard Cutout protects the
Calibrator and it cannot be changed (set by factory).
To set a Soft Cutout (see Figure 2-13):
1. Push .
2. Push .
3. Push (if security is set to high, enter password).
4. Push to select the Soft Cutout.
5. Push and to highlight a digit then push to increase the digit or push to
decrease the digit.
6. Push to save the Soft Cutout.