mA Input Calibration (-P Only)
Table 2-22 shows and describes the menu selections on the mA Input Calibration menu.
This menu is available on the -P Model only.
Table 2-22. mA Input Calibration Menu
66.03°C
mA INPUT CALIBRATION
mA C4: -0.0075
CAL
UUT
SYSTEM
MENU CALIB
mA C22: -0.0044
CAL mA
INPUT CAL DATE: 2012-05-01
MENU
1
2
3
gzs023.eps
Item Function
mA C4
[-P Only]
adjusts the measurement at 4 mA.
mA C22
[-P Only]
adjusts the measurement at 22 mA.
INPUT CAL DATE
[-P Only]
is the calibration date for the measurement function of the Calibrator.
System Information
Table 2-22 shows and describes the menu selections on the System Information menu.
Table 2-23. System Information Menu
66.03°C
SYSTEM INFORMATION
MODEL: 9190
SYSTEM
MENU
SYSTEM
INFO
SERIAL: 022
FW VER: 0.221/0.548
CAL DATE: 2012-05-01
INPUT CAL DATE: 01/05/2012
MENU
1
2
3
4
5
gzs024.eps
Item Function
MODEL shows the model number of the Calibrator.
SERIAL shows the serial number of the Calibrator.
FW VER shows the firmware version of the Calibrator.
CAL DATE shows the calibration date of the heat source.
INPUT CAL DATE
[-P Only]
shows the calibration date for the readout or the -P module.