Operation
Menus and Menu Navigation 2
2-27
Reference Probe Setup
Table 2-29 shows and describes the menu selections on the Reference Probe Setup menu.
Table 2-29. Reference Probe Setup Menu
66.03°C
REFERENCE PROBE SETUP
INPUT
SETUP
REF
INPUT
PROG
PROBE
SERIAL:
CAL DATE:
PROBE TYPE:
RTPW:
A: A:
B:
C:
A4:
B4:
AL
2000-01-01
ITS-90
100.0000
0.0000000 E-00
0.0000000 E-00
0.0000000 E-00
0.0000000 E-00
0.0000000 E-00
MENU
1
2
3
4
gzs031.eps
Item Function
SERIAL is the 10-digit, alpha-numeric serial number for the reference probe.
CAL DATE is the Calibration date for the reference probe.
PROBE TYPE is the probe conversion type to be set up. The type can be set to ITS-90, Callendar-
Van Dusen (CVD), IEC-751, or Resistance.
The PRT characterization coefficients that are shown depend on the probe type that was
selected. If ITS-90 is selected, the associated coefficients are a, b, c, a4, and b4. Coefficients a, b,
and c characterize the Reference Probe above 0 °C. Unused coefficients must be set to 0.
Note
If the probe connected is fitted with a smart connector, the PROGRAM prompt
appears when the Reference Probe Setup screen is exited. Select YES to update
the smart connector memory with the present probe setup or NO to leave the
memory unchanged.