Operation
Calibrator Setup 2
2-9
Caution
For safe operation and maintenance of the Product:
• Energize the Product for a 2-hour dry-out period before use, if the
Product was:
o In transport
o In a humid or semi-humid storage environment
o Not energized for more than 10 days
If the product is wet or has been in a wet environment, take
necessary measures to remove moisture prior to applying
power.
• Always operate this Product on a flat, level, stable surface.
• Do not store the Product at temperatures above 50 °C. The
Product has a refrigeration system and contains gasses
under pressure.
• Do not turn the Product upside down. The inserts will fall
out.
• To prevent damage to the cooling system, do not tilt the
Product on its side or upside down while the Product is
operating.
Connect to Mains Power
Use the 2-meter (6-foot) mains power cord to connect the Product to a 120 V ac or 230 V
ac outlet rated for at least 15 amps.
Turn On the Product
1. Push the “I” side of the power switch on the front panel of the Calibrator.
2. Monitor the Startup screen for errors while the product turns on. If an error shows,
contact Fluke Calibration.
Change Language
To change the display language:
1. Push .
2. Push .
3. Push .
4. Push .
5. Push or to highlight a language.
6. Push to set language.
Note
If the incorrect language is set by accident, push softkeys and at
the same time to temporarily switch back to the English language.
Set Display Contrast
With the Main screen shown in the display, push to increase or push to decrease
display contrast.