9190A
Operators Manual
2-26
Test UUT Calculation
Table 2-28 shows and describes the menu selections on the Test UUT Calculation menu.
Use this function to verify the accuracy of temperature calculations.
Table 2-28. Test UUT Calculation Menu
66.03°C
TEST UUT CALCULATION
INPUT
SETUP
SETUP
INPUT
TEST
CALC
SENSOR TYPE: TC
INPUT: 0.0000 mV
(TEMPERATURE: 0.000 °F)
MENU
1
2
3
gzs030.eps
Item Function
SENSOR TYPE selects the type of sensor for which the temperature calculation is to be tested.
INPUT sets the resistance value or thermocouple millivolt value for the probe connected.
TEMPERATURE shows the calculated temperature for the selected Sensor Type and resistance
or millivolt value.
REF Input Menu
The REF INPUT menu contains the parameters for the reference input to the readout
module of the Calibrator. The Reference Input is only compatible with PRTs with ITS-
90, Callendar-Van Dusen, or IEC-751 coefficients. Additionally, the Reference Input can
read straight resistance instead of temperature.
The probe serial number and coefficients can be found on the calibration certificate that
was shipped with the probe. If the probe requires calibration, contact an Authorized
Service Center to inquire about calibration services offered by Fluke Calibration (see
Chapter 1).